Flexiflo Quantum Pump Set with Piercing Pin and Flush Bag The product is used with Quantum pump to deliver ready-to-hang (RTH) product in enterally-fed patients.

Company
Abbott Laboratories
Recall Initiated
June 27, 2008
Posted
September 20, 2008
Terminated
September 19, 2012
Recall Number
Z-2305-2008
Quantity
57,120 units
Firm Location
Columbus, OH

Reason for Recall

Product was incorrectly labeled as "Top-Fill Enteral Nutrition Bag" instead of "Flexiflo Quantum Pump Set."

Distribution

Nationwide: AR, AZ, CA, CO, CT, FL, IA, ID, IL, IN, KS, KY, LA, MA, MI, MO, MS, NH, NJ, NV, NY, OH, OK, OR, PA, RI, TN, TX, VT, WA, WI and WY.

Lot / Code Info

List #50604 Lot #47822RY Use By DEC 2009

Root Cause

Labeling mix-ups

Action Taken

On June 27, 2008, the firm sent Important Device Recall letters to their customers. The letter instructed their customers to immediately inspect their inventory for the affected product, place it on hold, and report the information to back to them via account summary forms. Furthermore, if they are a distributor, they are required to contact their customers and notify them of this important device recall and supply them with the appropriate documentation to record the recall. Contact Abbott Nutrition, Abbott Laboratories at 1-614-624-3688 for assistance.