Terumo Advanced Perfusion System, 1 Roller Pump, 4 inch diameter; Catalog No. 801040, Terumo Cardiovascular Systems, Ann Arbor, MI 48103

Company
Terumo Cardiovascular Systems Corp
Recall Initiated
September 10, 2007
Posted
December 1, 2007
Terminated
February 23, 2008
Recall Number
Z-0321-2008
Quantity
16
Firm Location
Ann Arbor, MI

Reason for Recall

Failure to operate; The pump may fail to operate, and be difficult to hand crank, due to an internal electrical short condition.

Distribution

Worldwide Distribution-USA including states of Alabama, Arizona, Florida, Illinois, Indiana, Mississippi, Nebraska, North Carolina, Pennsylvania, and Virginia, and countries of Belgium, Hong Kong and United Arab Emirates.

Lot / Code Info

Serial Numbers 2110 through 2125.

Root Cause

Process control

Action Taken

Consignees were notified via an Urgent Medical Device Correction letter dated 9/10/07, which provided them guidance in the event the problem occurs and advised them they would be contacted to schedule replacement of the circuit boards.