ARCHITECT c8000 Processing Module, for In Vitro Diagnostics; List Number 106-01; Affected software: v2.10, v2.11, v2.12, v2.20, v2.20DB, v2.30, v2.60, v3.00 and v3.10; Distributed by Abbott Laboratories, Irving, TX.

This recall has been terminated (originally issued November 7, 2007).

Company
Abbott Laboratories, Inc
Recall Initiated
September 5, 2007
Posted
November 7, 2007
Terminated
October 4, 2010
Recall Number
Z-0146-2008
Quantity
2070 units.
Firm Location
Irving, TX
Official Source
View on FDA website ↗

Reason for Recall

Incorrect Sample/Test Identification: Due to a software timing defect with certain versions of ARCHITECT cSystem Software, test results may be incorrectly associated to a Sample Identification during specific sample-processing conditions of the Robotic Sample Handler.

Distribution

Worldwide Distribution: USA (including Puerto Rico), Argentina, Australia, Bahamas, Bermuda, Brazil, Cayman Islands, Canada, Chile, China, Colombia, Dominican Republic, Ecuador, El Salvador, Germany, Guatemala, Hong Kong, Jamaica, Mexico, New Zealand, Panama, Singapore, Thailand, Trinidad and Tobago, Uruguay, and Venezuela.

Lot / Code Info

All Serial Numbers: ARCHITECT cSystems using ARCHITECT System Software Versions: ---Software Version--List Number--Part Number-- 2.10-- 05F48-11-- 7-99077-01 2.11-- 05F48-12 --7-99077-02 2.12 --None --------7-99077-03 2.20 --05F48-13 ---7-201738-01 2.20DB --05F48-14 ---7-201738-02 2.30-- None --202246-01 2.60-- 05F48-17-- 7-203715-01 3.00-- 05F48-15-- 7-202246-01 3.10-- 05F48-18-- 7-203715-02

Root Cause

Software change control

Action Taken

Notification letters sent to consignees on 09/17/07. Consignees informed of the software defect and the conditions that would cause the software error to occur. Consignees told they would be contacted by an Abbott Field Service Rep who would install ARCHITECT System Software v3.11 and were asked to respond via fax using Customer Reply form acknowledging receipt and understanding of correction letter. Firm recalling to hospital/user level.