Bivona Adult Tracheostomy Tube, I.D. 8.5 mm, O.D 11.0 mm, Length 88 mm, TTS 11.8 mm, Mfrd by Smiths Medical Critical Care; Product code 670185.

This recall has been terminated (originally issued January 5, 2007).

Company
Smiths Medical Asd, Inc.
Recall Initiated
August 31, 2006
Posted
January 5, 2007
Terminated
May 4, 2007
Recall Number
Z-0339-2007
Quantity
197
Firm Location
Gary, IN
Official Source
View on FDA website ↗

Reason for Recall

Mislabeled- The outer diameter (OD) of the shaft is labeled as 11.0 m on the inner, outer and tray label in error. The OD is actually 11.8 mm.

Distribution

Worldwide Distribution ---- including USA states of Arkansas, California, Georgia, Illinois, Kansas, Minnesota, Missouri, Ohio, Pennsylvania, Texas and countries of Canada and Netherlands.

Lot / Code Info

Lots 1008187, 1023402 and 1012767.

Root Cause

Other

Action Taken

Distributors and hospital consignees were notified via recall letter dated 8/25/06 to return all unused product to the firm. Distributors were asked to notify their customers of the recall.