RecallDepth

DRLock Volar Plate Screw System, Model #DVR-900. The product is a kit contained in an aluminum tray case that contains the following: IMPLANTS: 6 stainless steel plates (2-lefts, 2-rights, and 1-right long and 1-right left), 48 distal locking-screws (ranging from 10-24mm), 32 distal locking pegs, 36 proximal locking screws, 24 proximal non-locking screws. INSTRUMENTS: 2 Bending Pliers*, 1 Bone Reduction Forcep*, 2 Hohmann Retractors *, 1 Ratchet Handle*, 1 Periosteal Elevator*, 1 Dingman Elevator *, 1 K & K Elevator *, four Driver tips *, five drill guides*, 2-1.6 Drill Bits *, 2-2.3 Drill Bits *, 1 Plate Holder Verbruge *, 2 Depth Gages *, 6 K wires.

This recall has been terminated (originally issued January 19, 2007).

Company
Orthohelix Surgical Designs Inc
Recall Initiated
August 3, 2006
Posted
January 19, 2007
Terminated
September 10, 2012
Recall Number
Z-0349-2007
Quantity
180
Firm Location
Akron, OH
Official Source
View on FDA website ↗

Reason for Recall

The 48 distal locking-screws (ranging from 10-24mm) contained in the kit are not within established specifications, specifically the minor diameter of the screws are undersized.

Distribution

The product was distributed to 5 consignees located in the U.S.: Archway Medical, Columbia, SC; Gulf Coast Surgical, Stafford, TX; Vaugham Medical, Wheat Ridge, CO; Belemy Medical, Cleveland, OH; and Crystal Clinic, Akron, OH.

Lot / Code Info

Tray kit (case) set number: ''BAA'', ''BAB'', ''BAC'', ''BAD'', ''BAE'', ''BAF'',  ''BAH'', and ''BAJ''. (The distal locking-screws contained in the kit are identified with lot numbers S104 and S200)

Root Cause

Other

Action Taken

The recalling firm contacted all consignees by phone on or about 8/3/06.

More recalls by Orthohelix Surgical Designs Inc

View all recalls by this company →