RecallDepth

Pronto Series Power Wheelchair, model M61.

Company
Invacare Corporation
Recall Initiated
May 2, 2005
Posted
June 17, 2005
Terminated
January 4, 2008
Recall Number
Z-0903-05
Quantity
137 devices.
Firm Location
Elyria, OH

Reason for Recall

If the arm of the wheelchair is overloaded, the joystick mounting brackett and/or arm pivot may fail due to variability in the plastic used by the supplier to make the components. This failure may cause the user or a caregiver to fall if they are leaning on the arm, and result in injury.

Distribution

The product was sent to consignees located in AL, AR, AZ, CA, CO, FL, GA, IA, IL, IN, KS, KY, LA, MD, ME, MI, MN, MO, MS, MT, NC, ND, NJ, NM, NV, NY, OH, OK, OR, PA, RI, SC, TN, TX, UT, VA, WA, WI, and WV. There was one foreign consignee in France: Poirer, Zac Papillon, 3720 Parcay Meslay, Corps, France

Lot / Code Info

All serial numbers between 05A000-05C999.

Root Cause

Other

Action Taken

The recalling firm contacted all consignees by phone between May 9 and May 25, 2005 and sent an urgent field correction needed letter, dated April, 2005 to all consignees.

More recalls by Invacare Corporation