Legend Lubricant System (Legend Gold, Gold Touch and Platinum motors and Legend Lubricant/Diffuser cartridge)

This recall has been terminated (originally issued January 9, 2003).

Company
Medtronics Midas Rex
Recall Initiated
November 11, 2002
Posted
January 9, 2003
Terminated
March 21, 2003
Recall Number
Z-0419-03
Quantity
PA100-94 cs, PM100-71ea, PM110-5 ea, PM200-24 ea
Firm Location
Fort Worth, TX
Official Source
View on FDA website ↗

Reason for Recall

Lubricant leaking from motor after sterilization causing bleed-through sterilization wrap and/or excess lubricant in instrument case.

Distribution

Nationwide: CA, OH, FL, PA, GA, AZ, LA, MI, TX, NJ, NM, NC, VA, WA, NE, AL, NY, TN

Lot / Code Info

LOT NUMBER (S) OR S/N(S):  PA100: T0262, T0367, T0369, T0418, T0466, T0545, T0547, T0657, T0659, T1114, T1784, T1785, T1937. PM100: T0474, T0476-T0479, T0481, T0482, T0486, T0487, T0490, T0491, T0494, T0495, T0498, T0502, T0505, T0506, T0508, T0709, T0710, T0765, T0769, T0776, T0980, T1122, T1124, T1133, T1134, T1137, T1139, T1140, T1142, T1298-T1300, T1303-T1305, T1310-T1312, T1363, T1374, T1377, T1391, T1568, T1574, T1575, T1578-T1580, T1586, T1587, T1590, T1594, T1595, T1597, T1601, T1603, T1684, T1687, T1688, T1695, T1699, T1702, T1708, T1710, T1711, T1722, T1739, T1755. PM110: T1014, T1622, T1624, T1628, T1631. PM200: T0453, T0614, T0615, T0617, T0620, T1149, T1326, T1332, T1430, T1528, T1530-T1532, T1535, T1552, T1556, T1561, T1562, T1564, T1637, T1638, T1640, T1648, T1655.

Root Cause

Other

Action Taken

Recall letter was sent on 11/11/2002 to all consignees requesting that the consignee contact the Medtronic Neurologic Technologies sales representative for replacement.