VARiS 1.4G and Vision/SomaVision Radiation Therapy Systems; VARiS versions 6.1.33 and 6.2.19 in conjunction with Eclipse 6.x/Soma Vision 6.x

This recall has been terminated (originally issued March 12, 2003).

Company
Varian Medical Systems, Inc.
Recall Initiated
January 31, 2003
Posted
March 12, 2003
Terminated
September 17, 2003
Recall Number
Z-0633-03
Quantity
12 units
Firm Location
Palo Alto, CA
Official Source
View on FDA website ↗

Reason for Recall

The product has a software problem. This could result in a treatment without a motorized wedge where one had been planned.

Distribution

Product was released for distribution to 7 hospitals in the US. Distribution was limited to the following States: CA, MA, NC, SC, PA, and LA. 1 (one) Canadian distribution includes Fraser Valley Cancer Center, Dept. of Radiation Oncology, 13750 96th Ave., Surrey, BC, V3V 1Z2, Canada. Foreign distribution includes Austria, Germany and Finland. The recall was appropriately extended to the user level; i.e., the hospitals/physicians who received the recalled product.

Lot / Code Info

Affected Serial Numbers: 44, 260, 315, 495, 746, 751, 873, 907, 984, 1098, 1131, and 9045

Root Cause

Software design

Action Taken

On 1/16/03 and 2/4/03 (2nd notification) the firm initiated the recall and their notification was via letters to inform its consignees about the new VARiS version 6.2.27 release.