RecallDepth

BacT/ALERT - FA Reagent Bottle The BacT/ALERT Microbial Detection System is used to determine if microorganisms are present in blood or other normally sterile body fluid samples taken from a patient suspected of having bacteremial/fungemia. The BacT/ALERT System and culture bottles provide both a microbial detection system and a culture media with suitable nutritional and environmental conditions for organisms commonly encountered in blood infections and other normally sterile body fluid infections. An inoculated bottle is placed into the instrument where it is incubated and continuously monitored for the presence of microorganisms that will grow in the BacT/ALERT FA bottle. BacT/ALERT FA Culture Bottles are used with the BacT/ALERT Microbial Detection System in qualitative procedures for enhanced recovery and detection of aerobic and facultative anaerobic microorganisms (bacterial and fungi) from blood and other normally sterile body fluids.

This recall has been terminated (originally issued May 24, 2011).

Company
Biomerieux, Inc.
Recall Initiated
April 13, 2011
Posted
May 24, 2011
Terminated
April 3, 2012
Recall Number
Z-2244-2011
Quantity
327,700 bottles
Firm Location
Durham, NC
Official Source
View on FDA website ↗

Reason for Recall

The charcoal suspension in a portion of the lot does not meet the intended density.

Distribution

Worldwide Distribution - USA (nationwide) and the countries of China, Tokyo, South Korea, Canada, Sweden, Austria, Poland, and Belgium

Lot / Code Info

Catalog number: 259791, Lot umber 1027481, Expiry: December 8, 2011

Root Cause

Other

Action Taken

BioMerieux sent an Urgent Product Removal Notice dated April 13, 2011, to all affected customers. The notice identified the product, the problem, and the action to be taken by the customer. Customers were instructed to: Evaluate positive results adhering to their institution's protocols for determining false versus true positives. Negative bottles should be checked by smear and/or subculture at some point prior to discarding as negative. Any remaining inventory of BacT/ALERT FA culture bottle lot 1027481 should be destroyed in accordance with their facility's policies and procedures. An Acknowledgement and Product Replacement Form was included to be completed and returned to bioMerieux or credit or replacement of unused bottles.

More recalls by Biomerieux, Inc.

View all recalls by this company →