Depuy Spine Mountaineer OCT Spinal System Screw Caddy Set, , Non Sterile Product Code: 2883-90-400 The MOUNTAINEER Screw caddy is intended to act as a holder in an implant set for implantable screws prior to use in surgery. Further, the caddy provides a means to organize the screws in accordance with the length of the screws.

Company
Depuy Spine, Inc.
Recall Initiated
October 2, 2010
Posted
December 27, 2010
Terminated
April 15, 2011
Recall Number
Z-0818-2011
Quantity
121 units
Firm Location
Raynham, MA

Reason for Recall

The scale which serves as a secondary means of identifying screw length is incorrectly etched on the screw caddy

Distribution

Nationwide distribution: USA including states of: AR, AZ, CA, FL, HI, IA, IL, KY, LA, MA, MI, MN, MO, NC, NY OH, PA, TN, TX, VA, and WA.

Lot / Code Info

Not lot coded

Root Cause

Labeling design

Action Taken

Depuy Spine issued an "URGENT-VOLUNTARY PRODUCT RECALL NOTIFICATION" via e-mail on October 2, 2010 to customers. The notification describes the product, problem and action to be taken by customers. The customers were instructed to please distribute this information to appropriate personnel at their facility and return any caddies that they may have to their local distributor for DePuy Spine products. Depuy will return devices to customer promptly (within two business days), after reworking the devices by removing the scale altogether. Once the product is fully corrected, Depuy will provide the customers with a new and updated unit with the corrected scale markings. If you should have any questions or concerns regarding this action, please contact DePuy Spine Customers Service at 800-227-6633.

More recalls by Depuy Spine, Inc.

View all recalls by this company →