Vacuette 2 ml Lithium Heparin, Collection Tube, Manufactured by: Greiner-Bio One North America, Inc., 4238 Capital Drive, Monroe, NC. Venous blood collection tubes.

Company
Greiner Bio-One North America, Inc.
Recall Initiated
August 25, 2010
Posted
December 16, 2010
Terminated
May 15, 2012
Recall Number
Z-0623-2011
Quantity
18 cases (total 21,600 pieces)
Firm Location
Monroe, NC

Reason for Recall

Lithium heparin solution was contaminated by EDTA K (potassium) solution due to an operator error and a misinterpretation of a test result. Contaminated tubes will produce inaccurate test results.

Distribution

Nationwide Distribution: USA, including the states of CA, FL, IL, IN, KS, KY, ND, NE, NH, OH, SC, SD and WI.

Lot / Code Info

Item number: 454237, Lot number: B061006

Root Cause

Other

Action Taken

Greiner Bio-One instructed customers to stop distribution and to complete an attached Product Disposition form and return it via fax to Greiner Bio-One. They were further instructed to destroy any product they may have. Distributors were instructed provide a list of their customers that have received the affected item via Excel file so that they can be notified of the recall.