Pioneer Plus Catheter PPlus 120 dual lumen device, inserted through a 6F introducer sheath and placed percutaneously into a peripheral vessel. The device tracks to its intended site in the vasculature over a standard length, commercially available 0.014" (0.36 mm) Rapid Exchange (RX) tracking guide wire. It utilizes an extendable, hollow Nitinol guide tip (needle) to facilitate the redirection and placement of a 0.014 (0.36 mm) Over the Wire (OTW) guide wire into peripheral vessels. Specification Developer: Medtronic, Minneapolis, MN. Manufactured by: Accellent, Laconia NH The Pioneer Plus Catheter is intended to facilitate placement and positioning of catheters within the peripheral vasculature. The Pioneer Plus Catheter also provides an intra-luminal cross-sectional ultrasound image of the area of interest to facilitate placement of guide wires beyond stenotic lesions (e.g., sub-total, total or chronic total occlusions) prior to additional intervention (i.e. PTCA, stent, etc.). The Pioneer Plus Catheter is not indicated for use in the coronary or cerebral vasculature.
This recall has been terminated (originally issued January 6, 2011).
- Company
- Medtronic Cardiovascular
- Recall Initiated
- April 30, 2010
- Posted
- January 6, 2011
- Terminated
- January 6, 2011
- Recall Number
- Z-0864-2011
- Quantity
- 1,080 units
- Firm Location
- Santa Rosa, CA
- Official Source
- View on FDA website ↗
Reason for Recall
Needles may fail to retract and thus cause injury.
Distribution
Worldwide Distribution - USA, Germany, Netherlands, Slovakia, and Spain
Lot / Code Info
All Lots, no serial numbers.
Root Cause
Nonconforming Material/Component
Action Taken
Medtronic sent an Urgent Medical Device Safety Information letter to all affected customers. The letter identified the product, the problem,and the action to be taken by the customer. Sales Representatives were to contact each account and complete an acknowledgement form indicating the Physicians had been notified. For questions regarding this recall customers were to contact their local representative.
More recalls by Medtronic Cardiovascular
Medtronic Racer Biliary Stent System, Model number XD518YF, UPC code 00721902758164 in a box contain...
Jun 14, 2010
Complete SE Biliary Stent System, SC680L (6 mm x 120 mm), lot number 0000983806, UPC code 0061339941...
Jan 19, 2010
Complete SE Biliary Stent System, SC5120L (5 mm x 12 mm), lot number 0000983799, UPC code 0061399418...
Jan 19, 2010