RX ACCULINK Carotid Stent System 10 X 20mm, Part Number: 1011342-20, Lot Number: 0072061 The RX Acculink Carotid Stent System, used in conjunction with Abbott Vascular RX Accunet embolic protection system, is indicated for the treatment of patients at high risk for adverse events from carotid endarterectomy who require carotid revascularization and meet the criteria outlined below. (1) Patients with neurological symptoms and >50% stenosis of the common or internal carotid artery by ultrasound or angiogram OR patients without neurological symptoms and >80% stenosis of the common or internal carotid artery by ultrasound or angiogram, AND (2) Patients must have a reference vessel diameter within the range of 4.0 mm and 9.0 mm at the target lesion.

Company
Abbott Vascular
Recall Initiated
September 10, 2010
Posted
November 10, 2010
Terminated
November 16, 2010
Recall Number
Z-0338-2011
Quantity
42 units total for all products
Firm Location
Temecula, CA

Reason for Recall

The recall was initiated because Abbott vascular has discovered through internal testing that the affected lots may not meet the firm's quality specifications for catheter shaft tensile strength. If the affected device is used it may result in acute stent deployment difficulties and subsequent intervention.

Distribution

Nationwide Distribution including AL, CA, CT, DC, FL, GA, IA, IL, IN, MA, ME, MI, MN, MO, OH, OK PA, TX, VA, and WI

Lot / Code Info

Lot Number: 0072061

Root Cause

Pending

Action Taken

Abbott Vascular sent an URGENT DEVICE RECALL letter dated September 10, 2010, to all consignees. The letter identified the product, the problem, and the action to be taken. The Sales Representatives were to personally visit each account to deliver the recall letter and assist in identifying and returning any unused devices to Abbott Vascular. Also, Abbott Vascular Regulatory Compliance initiated contact with the Risk Management department at each account by phone to verify their mailing address and provide a courtesy copy of the recall letter. The consignees were instructed to work with their local account representative to review their inventory, complete the Recall Effectiveness Check and return identified products to Abbott Vascular. For questions regarding this recall call (951) 914-3324.