Freestyle Navigator Continuous Glucose Monitor System, ART 16467 Rev A, ABB System Kit MG, EMEA, DE. Abbott Item Number 70965-01, distributed by Abbott Diabetes Care Inc, Alameda, CA. Medical device indicated for continually recording interstitial fluid glucose levels in people ages 18+ with diabetes mellitus for improving diabetes management. The device is indicated to be used in hospital and home settings to aid health care.
- Company
- Abbott Diabetes Care, Inc.
- Recall Initiated
- April 14, 2009
- Posted
- April 30, 2010
- Terminated
- March 31, 2011
- Recall Number
- Z-1448-2010
- Quantity
- 5,449 of total units
- Firm Location
- Alameda, CA
Reason for Recall
Potential failure or inaccurate readings--the device's plastic housing near the battery compartment may crack, and allow moisture to enter.
Distribution
Worldwide Distribution -- United States (OH, FL, NJ, CA, MS, MN, MA, and NY), France, Australia, the Netherlands, Sweden, the UK, Israel, Ireland and Germany.
Lot / Code Info
Abbott Item # 70965-01; with multiple lot numbers, kits serial numbers, and Tx serial numbers. Lot numbers are these: 01K2578, 01K2638, 01K2698, 01K3238, 01K3328, 01K3338, 01K3488, 01M0108, 01M0238, 01M0368, 01M0798, 01M0848, 01M1768, 01M2048, 01M2138 AND 02M0238. Kit Serial numbers beginning with: BAAH 156 or BAAH 200 or BAAH 207 or BAAH 220 or BAAH 223 or BAAH 243 or BAAH 261or BAAH 299 or BAAH 341 and followed by a five (5 )digit number. Tx Serial numbers beginning with: AAAH 090 or AAAH 139 or AAAH 158 or AAAH 187 or AAAH 195 or AAAH 212 or AAAH 221 or AAAH 228 or AAAH 239 or AAAH 268 or AAAH 283 or AAAH 296 and followed by a five (5) digit number.
Root Cause
Process control
Action Taken
Notification letters were issued to FreeStyle Navigator customers via certified mail and notification was posted on the Abbott Diabetes Care's Website. Customers were notified of the potential issue and actions to take: 1) if moisture is observed in the FreeStyle Navigator transmitter now or at any time, discontinue use of the transmitter and call Abbott Diabetes Care Customer Care at 1-800-418-6427 and 2) if a loss of connection is encountered or if your results are not reliable or are inconsistent with how you feel, consult your User's Guide for troubleshooting guidelines. If the issue is not resolved, call Abbott Diabetes Care Customer Care at 1-800-418-6421. For outside-of-US accounts, the firm's regional offices will distribute notices to identified customers in the appropriate language.