BD Gram Stain Kit, containing 1- 250 mL bottle of Gram Crystal Violet, and additional reagents: Gram Iodine (Stabilized), Gram Decolorizer, Gram Safranin. Gram Stain Kits and Reagents are used to stain microorganisms from cultures or specimens by the differential Gram method.

This recall has been terminated (originally issued July 18, 2012).

Company
Becton Dickinson & Co. Bd Diagnostic Systems
Recall Initiated
April 12, 2012
Posted
July 18, 2012
Terminated
December 10, 2012
Recall Number
Z-2031-2012
Quantity
539 kits
Firm Location
Sparks, MD
Official Source
View on FDA website ↗

Reason for Recall

Potential for increase level of artifacts to appear as large, spherical and easy to distinguish or small & more difficult to distinguish from gram positive cocci.

Distribution

Nationwide Distribution.

Lot / Code Info

Lot 1348873, Exp April 30, 2013

Root Cause

Material/Component Contamination

Action Taken

BD Diagnostic Systems sent an "URGENT PRODUCT RECALL" letter dated April 2012 to all affected customers. The letter identifies the product, problem, and actions to be taken by the customers. The letter instructs customers to discontinue use and discard the affected product. An Acknowledgement Form was attached for customers to complete and return via fax to 410-316-4258. Contact the BD Customer Service Department for questions regarding this recall.

More recalls by Becton Dickinson & Co. Bd Diagnostic Systems