Philips Easy Diagnost Classic-Stationary x-ray system Catalog Numbers: 70642, 70643, 706030, 706031, 706032 Stationary x-ray system, Universal RF system for general use.

This recall has been terminated (originally issued August 15, 2012).

Company
Philips Healthcare Inc.
Recall Initiated
July 11, 2012
Posted
August 15, 2012
Terminated
January 23, 2017
Recall Number
Z-2208-2012
Quantity
12 units
Firm Location
Andover, MA
Official Source
View on FDA website ↗

Reason for Recall

Inclination switch in the stand may have been mounted incorrectly, as a result, the switch could not stop the tilting movement of the table base due tol hardware or software errors.

Distribution

Worldwide Distribution - USA including IN, MA, MI, MO, NC, NY PA, SC, and WI and to the countries of Canada, HongKong, New Zealand.

Lot / Code Info

Serial Numbers: SN07000141 SN07000109 SN5000162 SN5000421 SN07000667 200148 200115 200081 300114

Root Cause

Equipment maintenance

Action Taken

Philips issued an Urgent Field Safety Notice dated July 9, 2012, to notify users that the inclination swtich may be mounted incorrectly. A Field Safety Engineer will be dispatched to verify if a switch was installed properly, and in the event it was not installed properly they will re-install it properly. The Field Service Engineer will add a label as described in the spare part replacement manual, which is part of the FCO kit.. Customers were instructed review the information contained in the Field Safety Notice and to retain a copy of the notice with the equipment Instruction for Use. For further questions customers were instructed to call their Phillips representataive Technical Support Line at 1-800-722-9377. For questions regarding this recall call 978-687-1501