Stryker Xia 3 Anti-Torque Key Manual Surgical Instrument; Catalog Number 48237026. Product Usage: : Reusable instrument used to aid in the final tightening of the spinal construct. Intended Use: The Anti-Torque Key allows the Torque Wrench to align with the tightening axis and maximizing the torque needed to lock the implant assembly during final tightening of the spinal construct.

Company
Stryker Spine
Recall Initiated
January 12, 2012
Posted
June 5, 2012
Terminated
March 21, 2013
Recall Number
Z-1706-2012
Quantity
95 units
Firm Location
Allendale, NJ

Reason for Recall

Stryker received four reports regarding a handle which separates at the point where it is welded to the shaft on the Xia 3 Anti-Torque Key during surgery.

Distribution

US Nationwide Distribution

Lot / Code Info

Lot Codes (510k exempt): 07F116, 07F117, 085525, 086577, 086578, 087232, 087233, 088604.

Root Cause

Mixed-up of materials/components

Action Taken

Stryker Spine sent an Urgent Product Recall letter dated January 12, 2012 with an attached Customer Response forms via FEDEX to affected customers. The letter identified the affected product, problem and actions to be taken. Customers were instructed to examine their inventory to identify the affected product and return using the prepaid mailing label provided. The letter asked customers to complete the Customer Response Form and fax to 201.760.8370. For questions call 201.760.8298.