Talent Captivia Stent Graft System - FreeFlow configurations only. Model numbers affected have the following designation: TFXXXXXXXXCP and TBXXXXXXXXCP. Medtronic Inc. Santa Rosa, CA 95403. The Talent Thoracic Stent Graft System is designed to treat thoracic aortic aneurysms using an endovascular approach. When placed within the target lesion, the stent graft provides an alternative conduit for blood flow within a patients vasculature by excluding the lesion from blood flow and pressure.

Company
Medtronic Vascular, Inc.
Recall Initiated
February 21, 2012
Posted
March 23, 2012
Terminated
July 12, 2012
Recall Number
Z-1284-2012
Quantity
1523 units
Firm Location
Santa Rosa, CA

Reason for Recall

Complaints received related to the inability/difficulty to release the proximal (FreeFlow) bare stent using the tip capture mechanism.

Distribution

Nationwide Distribution

Lot / Code Info

TB2222C116CP, TF2424C116CP, TF2626C116CP, TF2828C116CP, TF3030C115CP, TF3232C115CP, TF3434C115CP, TF3636C114CP, TF3838C114CP, TF4040C114CP, TF4242C113CP, TF4444C113CP,  TF4646C112CP, TB2222C161CP, TF2424C161CP, TF2626C161CP, TF2828C161CP, TF3030C160CP, TF3232C160CP, TF3434C160CP, TF3636C159CP; TF3838C159CP, TF4040C159CP, TF4242C158CP, TF4444C158CP, TF4646C157CP, TB2222C201CP, TF2424C201CP, TF2626C201CP, TF2828C201CP, TF3030C200CP, TF3232C200CP, TF3434C200CP, TF3636C199CP, TF3838C199CP, TF4040C199CP, TF4242C198CP, TF4444C198CP, TF4646C197CP.

Root Cause

Process design

Action Taken

Medtronic, Inc. sent an Urgent Medical Device Safety Information letter dated February 2012, to all affected customers. The letter identified the product, the problem, and the corrective action taken by Medtronic. Medtronic updated the Instructions for Use (IFU) manual. For questions regarding this recall call 707-566-1110.