GALT VALVED TEARAWAY INTRODUCER SET GALT VTI RX ONLY UNITS 7 STERILE EO Product Usage: Used in percutaneous procedures to introduce catheters and other intravascular devices into the peripheral vasculature.

This recall has been terminated (originally issued September 4, 2014).

Company
Galt Medical Corporation
Recall Initiated
July 25, 2014
Posted
September 4, 2014
Terminated
February 26, 2016
Recall Number
Z-2588-2014
Quantity
11,155 units in total
Firm Location
Garland, TX
Official Source
View on FDA website ↗

Reason for Recall

The adhesive used in the product could dislodge when objects are inserted through the valve which could pose an embolic risk to the patient.

Distribution

Worldwide Distribution - US (nationwide) Canada, Germany, the Netherlands, Ireland, Switzerland, France, Italy, and Greece.

Lot / Code Info

Catalog Numbers: NT-104-06, INT-104-07, INT-104-08, INT-104-09, INT-104-10, INT-104-11, INT-104-12, INT-104-17, INT-104-18, INT-104-19, INT-104-20, INT-104-21, INT-104-22, INT-104-23, INT-108-06, INT-108-07, INT-108-09, INT-108-10, INT-108-11, INT-108-12, INT-108-17, INT-108-18, INT-108-19, INT-108-20, INT-108-21, INT-108-22, INT-108-22, VTI-002-06, VTI-002-07, VTI-002-08, VTI-002-09, VTI-002-09, VTI-002-10, VTI-002-11, VTI-002-12, VTI-002-17, VTI-002-18, VTI-002-19, VTI-002-20, VTI-002-21, VTI-002-22, VTI-002-23, VTI-003-06, VTI-003-07, VTI-003-08, VTI-003-09, VTI-003-10, VTI-003-11, VTI-003-12, VTI-003-17, VTI-003-18, VTI-003-19, VTI-003-20, VTI-003-21, VTI-003-22, and VTI-003-23.

Root Cause

Nonconforming Material/Component

Action Taken

Galt Medical sent an Urgent Medical Device Recall letters to all affected customers. The letter identified the affected product, reason for recall and actions to be taken. Customers were asked to identify and segregate the recalled lot (s) that is in their possession, complete and return the enclosed Recall Reply form, and return the recalled product to Galt Medical Corp. For questions contact your local sales representative or Recall Coordinator at 214-778-1306.