Roche cobas IT 1000 Product Usage: Intended Use: An electronic device intended to store, retrieve, and process laboratory data.

This recall has been terminated (originally issued September 1, 2014).

Company
Roche Diagnostics Operations, Inc.
Recall Initiated
July 31, 2014
Posted
September 1, 2014
Terminated
June 16, 2015
Recall Number
Z-2563-2014
Quantity
148 Units
Firm Location
Indianapolis, IN
Official Source
View on FDA website ↗

Reason for Recall

Potential that test results would be assigned to the wrong Patient ID if cobas IT 1000 receives an ADT ADMIT, TRANSFER or MERGE event message that contains a genuine Patient ID (e.g. 999999), which is the same as what has been configured as the Generic Patient ID (999999 by default).

Distribution

US Nationwide Distribution in the states of: AK, AL, AR, AZ, CA, CT, DE, FL, GA, HI, IA, IL, IN, KS, KY, LA, MD, ME, MI, MO, NC, NE, NH, NJ, NY, OH, OR, PA, PR, SC, TN, TX, VA, VT, WA, WI, and WV.

Lot / Code Info

Part Numbers: 04420306001 Serial #'s: 2706, 2724, 2743, 2744, 2745, 2749, 2750, 2751, 2752, 2754, 2755, 2764, 2770, 2771, 2772, 2773, 2774, 2775, 2776, 2780, 2781, 2782, 2783, 2784, 2785, 2789, 2790, 2791, 2792, 2793, 2794, 2795, 2796, 2797, 2798, 2799, 2808, 2814, 2815, 2816, 2818, 2819, 2823, 2824, 2825, 2826, 2828, 2829, 2830, 2831, 2847, 2848, 2849, 2857, 2858, 2859, 2860, 2861, 2862, 2863, 2864, 2865, 2904, 2905, 2906, 2908, 2909, 2914, 2915, 2916, 2917, 2920, 2921, 2928, 2931, 2932, 2933, 2934, 2935, 2936, 2937, 2938, 2939, 2940, 2941, 2943, 2983, 2984, 2985, 2986, 2987, 2990, 2991, 2992, 2993, 2994, 2995, 2996, 2997, 2998, 3006, 3007, 3008, 3009, 3016, 3020, 3021, 3023, 3024, 3025, 3026, 3027, 3048, 3049, 3050, 3051, 3053, 3057, 3058, 3059, 3060, 3061, 3065, 3066, 3067, 3068, 3069, 3070, 3071, 3072, 3073, 3074, 3075, 3076, 3077, 3079, 3082, 3083, 3084, 3085, 3086, 3087, 3089, 3090, 3104, 3105, 3106, and 3107.

Root Cause

Software design

Action Taken

Roche sent an Urgent Medical Device Correction letter dated July 31, 2014 via UPS Ground (receipt required) to affected customers. The letter identified the affected product, problem and actions to be taken. Customers were instructed to follow the steps provided in the letter to prevent assigning patient results to an incorrect generic patient ID. Please contact the Roche Support Network Customer Support Center at 1-800-440-3638, prompt 5, 24 hours a day, seven days a week if you have questions about the information contained in this UMDC.