RecallDepth

Treatment packs used on the VisuMax Laser Keratome. Size M Manufactured by Carl Zeiss Meditec AG, Jena, Germany. The VisuMax Laser Keratome is indicated for the following: In the creation of a corneal flap in patients undergoing LASIK surgery or other treatment requiring lamellar resection of the cornea. Patients undergoing surgery or other treatment requiring initial lamellar resection of the cornea. In the creation of a lamellar cut/resection of the cornea for lamellar keratoplasty. In the creation of a cut/incision for penetrating keratoplasty and corneal harvesting.

This recall has been terminated (originally issued March 7, 2014).

Company
Carl Zeiss Meditec Ag Carl Zeiss Promenade 10 Jena Germany
Recall Initiated
February 3, 2014
Posted
March 7, 2014
Terminated
March 7, 2014
Recall Number
Z-1153-2014
Quantity
1 package containing 10 individual treatment packs
Official Source
View on FDA website ↗

Reason for Recall

Carl Zeiss Treatment pack, lot number M 130010, size M, may actually contain size S.

Distribution

US Distribution: NV only.

Lot / Code Info

P/N 1462-333, Size M, Lot # M130010

Root Cause

Mixed-up of materials/components

Action Taken

The firm, Zeiss, sent an "Important Field Safety Advisory Note" letter dated February 3, 2014 to its customer. Letter indicates that a phone call preceded the letter on January 23, 2014. The letter also describes the product, problem and actions to be taken. The customers were instructed to quarantine all treatment packs size "M", batch number M 130010; return the material to Carl Zeiss Meditec, Inc., 5160 Hacienda Dr, Dublin, CA 94568 and Zeiss will issue a credit for all material received, upon receipt of the returned materia, and to complete and return the Conformation Form with the envelop provided. If you have any questions regarding this field corrective action, please call 925-557-4832 or email at paul.shahan@zeiss.com.

More recalls by Carl Zeiss Meditec Ag Carl Zeiss Promenade 10 Jena Germany

View all recalls by this company →