Siemens Artis Q and Artis Q biplane systems Product Usage: Interventional Fluoroscopic X-ray

This recall has been terminated (originally issued March 27, 2014).

Company
Siemens Medical Solutions Usa, Inc
Recall Initiated
December 5, 2013
Posted
March 27, 2014
Terminated
November 6, 2014
Recall Number
Z-1202-2014
Quantity
19
Firm Location
Malvern, PA
Official Source
View on FDA website ↗

Reason for Recall

There exists the possibility that during clinical operation, a breakdown of the x-ray tube could result in loss of x-ray imaging for the Artis Q and Artis Q biplane systems with a flat detector as 40 HDR (large 30x40).

Distribution

USA Nationwide Distribution in the states of TN, MA, CA, IL, OH, FL, UT, WA, GA, IA, NY, MO, IN, and CO.

Lot / Code Info

model numbers 10848280, 10848281, 10848282, and 10848283.

Root Cause

Device Design

Action Taken

Siemens sent an Important Customer Safety Notice letter dated November 22, 2013, to affected customers. The letter identifed the affected product, problem and actions to be taken. Customers were instructed to discontinue use until the affected product has been successfully replaced by a Siemen's representative.at no cost.