RecallDepth

PowerPort Slim Implantable Port, indicated for patient therapies requiring repeated access to the vascular system. Product Codes: 1716000, 1716070

This recall has been terminated (originally issued March 27, 2014).

Company
Bard Access Systems
Recall Initiated
February 27, 2014
Posted
March 27, 2014
Terminated
September 5, 2014
Recall Number
Z-1300-2014
Quantity
777
Firm Location
West Salt Lake City, UT
Official Source
View on FDA website ↗

Reason for Recall

Labeling discrepancy.The label states that the implantable port contains silicone suture plugs, however approximately 23% of the units contained in the affected lots have a PowerPort Slim titanium implantable port without silicone suture plugs.

Distribution

US Nationwide distribution including WA, KY, IL, and IA.

Lot / Code Info

Lot Numbers: REXA1302, REXA1350, REXB1160

Root Cause

Packaging

Action Taken

Notified by Urgent Customer Notification Letter date February 13, 2014, customers were asked to immediately examine their inventory and identify any product subject to this notification. In addition, if they have further distributed this product, please identify customers and notify them at once of this notification. Customers may include a copy of this letter in your notification to customers. Customers may choose to use the involved product with the understanding that the kit may contain a PowerPort Slim implantable port without suture plugs, or may return the applicable product for replacement or credit.