5/10/12 mm Weck Vista" Optical Bladeless Laparoscopic Access Port, Product Usage: An endoscope and accessories is a device used to provide access, illumination, and allow observation or manipulation of body cavities, hollow organs, and canals.

This recall has been terminated (originally issued March 25, 2014).

Company
Teleflex Medical
Recall Initiated
March 11, 2014
Posted
March 25, 2014
Terminated
August 15, 2017
Recall Number
Z-1283-2014
Quantity
55,521 ea. in total
Firm Location
Research Triangle Park, NC
Official Source
View on FDA website ↗

Reason for Recall

Complaints were received of leakage of insufflation gas through the device, which can cause difficulty or delay in the completion of laparoscopic surgical procedures.

Distribution

Worldwide Distribution - USA Nationwide including Puerto Rica and the countries of: Australia, Brazil, China, Canada, Malaysia, Korea, Ireland Singapore and Vietnam.

Lot / Code Info

Product Code: 405910RC, Lot Numbers: 01J1200010, 01J1200242, 01K1200070, 01A1300141, 01B1300052, and 01D1300319.

Root Cause

Nonconforming Material/Component

Action Taken

Teleflex sent an Urgent Medical Device Recall Notification letter dated March 11, 2014 to all affected customers. The letter identified the affected product, problem and actions to be taken. Customers were instructed to return all affected product to Teleflex Medical per the instructions provided in the letter. Customers were asked to complete the enclosed Recall Acknowledgment Form and fax it to 1-866-804-9881, Attn: Customer Service. For questions contact your local sales representative or Customer Service at 1-866-246-6990.