Philips HeartStart MRx Monitor/Defibrillator with CO2/EtCO2 Measurement The HeartStart MRx is for use for the termination of ventricular tachycardia and ventricular fibrillation. The device is for use by qualified medical personnel trained in the operation of the device and qualified by training in basic life support, advanced cardiac support, or defibrillation. It must be used by or on the order of a physician. The HeartStart MRx EtCO2 Option monitors carbon dioxide (CO2) and measures end-tidal carbon dioxide (EtCO2). The option provides an EtCO2 value and a CO2 waveform on the devices display.

This recall has been terminated (originally issued December 24, 2013).

Company
Philips Medical Systems, Inc.
Recall Initiated
December 10, 2013
Posted
December 24, 2013
Terminated
April 18, 2017
Recall Number
Z-0570-2014
Quantity
18,529 units
Firm Location
Andover, MA
Official Source
View on FDA website ↗

Reason for Recall

MRx defibrillator displays a -?- for EtCO2 and does not display EtCO2 values when patient CO2 level falls below 7.6 mmHg

Distribution

US Nationwide and OUS.

Lot / Code Info

Software versions: F.02, R.02, 9.xx or any earlier versions   Addendum Models: M3532A, M3535A, M3536AM, M3536MC, M3536M2, M3536M4, M3536M5, M3536M6 M3535A- HeartStart MRx Defibrillator Monitor M3536A- HeartStart MRx Defibrillator Monitor EtC02 Options A04 A05 A06 All M3532A Upgrade

Root Cause

Software design

Action Taken

Philips Healthcare issued an Customer Information letter dated February 20, 2012, to all affected customers. The letter identified the product the problem and the action needed to be taken by the customer with the addendum to the Instructions for Use. If you have a ? in the parameter block and no CO2 waveform on the display, the waveform source is invalid. If invalid, check patient, confirm airway status and examine the FilterLine to make sure it is connected to the device and not kinked or pinched. For more information on the EtCO2 Option see Monitoring Carbon Dioxide chapter in the HeartStart MRx Instructions for Use. Should you have any questions or concerns, please contact your local Philips representative at 1-800-722-9377.