Cordis ADROIT 6F Guiding Catheter Product Usage: For intravascular introduction of interventional and/or diagnostic devices into the coronary or peripheral vascular system.

This recall has been terminated (originally issued October 1, 2013).

Company
Cordis Corporation
Recall Initiated
August 12, 2013
Posted
October 1, 2013
Terminated
September 30, 2014
Recall Number
Z-0003-2014
Quantity
1238 units
Firm Location
Miami Lakes, FL
Official Source
View on FDA website ↗

Reason for Recall

Due to a labeling error related to the conversion of the inner diameter of the catheter from inches to millimeters. The English dimension is correct and is the primary dimension.

Distribution

Worldwide Distribution - USA Nationwide in the states of AZ, CA, FL, ID, LA, MA, MD, MI, MN, MO, NC, NH, NY, OH, OK, PA, SC, TX, VA, WI, and WV and the countries of: Austria, Belgium, Czech Republic, France, Germany, Greece,Hungary, Israel, Italy, Latvia, Netherlands, Poland, Portugal, Russian Federation, Slovenia, Spain, Sweden, Switzerland, United Arab, and United Kingdom

Lot / Code Info

Catalog Lots  ---------------------------------- 67200000 15918614 67200200 15892736 67200400 15892737, 15897416 67200500 15901314 67203400 15922021 67203600 15897418 67204000 15925687 67205200 15892738, 15892739 67205400 15885645, 15887472, 15897420 67205600 15901319 67206000 15897422 67207200 15892740 67208000 15892741 67208200 15887473, 15887474, 15897424 67208300 15901321 67212600 15916942 67213000 15916943 67227000 15901323

Root Cause

Error in labeling

Action Taken

Cordis sent an Urgent Medical Device Recall letter dated August 12, 2013 Acknowledgement Form to all affected customers. The letter identifed the affected product, problem and actions to be taken. Customers were instructed to immediately identify and set aside all affected product in a manner that ensures the product will not be used. Customers were asked to review, complete, sign and return the enclosed Acknowledgement form directly to Cordis to the fax number on the form. For questions call the Customer Service number at 800-561-7683.