STERRAD 100NX Cassettes The STERRAD Sterilization System is a low temperature general purpose sterilizer used to sterilize heat and moisture sensitive reusable medical devices.

This recall has been terminated (originally issued October 22, 2013).

Company
Advanced Sterilization Products
Recall Initiated
September 18, 2013
Posted
October 22, 2013
Terminated
January 13, 2014
Recall Number
Z-0044-2014
Quantity
2,995 units
Firm Location
Irvine, CA
Official Source
View on FDA website ↗

Reason for Recall

Advanced Sterilization Products (ASP) is recalling the three lots (13F049, 13F050, 13F051) STERRAD 100NX Cassettes because ASP has determined that the leak indicator on the cassettes is effective for only 12 months of the cassette's labeled 15-month shelf-life.

Distribution

Nationwide Distribution

Lot / Code Info

Lot# 13F050, 13F051, 13F049

Root Cause

Under Investigation by firm

Action Taken

Advanced Sterilization Products (ASP) sent an Urgent Medical Device Recall letter on September 18, 2013, to all affected customers. The letter informed the customers of the problems identified and the actions to be taken. Customers were instructed to examine their inventory and return all affected cassettes using the enclosed prepaid UPS return lave to Stericycle. Customers were asked to count their inventory and record the data on the enclosed Business Reply Card and Packing Slip. Customers were asked to provide the notice to anyone in their facility that needs to be informed. Customers were also asked to maintain a copy of the notice with the affected product. Customers with questions or suspected problems were instructed to contact Stericycle directly at 1-877-257-7119. For questions regarding this recall call 949-453-6400.