Medical Action Industries, Venous Access Tray, using BD Q-Syte Luer Access Split Septum (BD #385100)), Part number: 69498.

Company
Medical Action Industries Inc
Recall Initiated
November 6, 2009
Posted
March 30, 2010
Terminated
June 1, 2011
Recall Number
Z-1074-2010
Quantity
4,280 kits
Firm Location
Arden, NC

Reason for Recall

Manufacturing defect may allow air entry into a central venous catheter, possibly resulting in an air embolism.

Distribution

Nationwide distribution

Lot / Code Info

Part number: 69498, Lot numbers: 62282, 69902, 70350, 70854 and 72108.

Root Cause

Nonconforming Material/Component

Action Taken

On 11/6/2009 a letter was drafted and sent by email to three (3) Medical Action Industries customers. They were instructed to return the recalled products in inventory to MAI. They further advised, if the product was distributed to notify their customer, inform them of the recall and instruct them to return the affected lots to MAI.