Tibial Array and Femoral Array. Model Number(s): 0150ARR00001 and 0150ARR00002. Manufactured by: MAKO Surgical, Corp. The devices are accessories used with MAKO's Tactile Guidance System (TGS). The Tactile Guidance System is intended to assist the surgeon in providing software defined spatial boundaries for orientation and reference information to anatomical structures during orthopedic procedures.

This recall has been terminated (originally issued February 11, 2009).

Company
Mako Surgical Corporation
Recall Initiated
May 2, 2008
Posted
February 11, 2009
Terminated
August 21, 2009
Recall Number
Z-0915-2009
Quantity
44
Firm Location
Davie, FL
Official Source
View on FDA website ↗

Reason for Recall

The product in its current state may fail during use, which would cause the instrument to be in-accurate.

Distribution

Nationwide Distribution USA including states of FL, NY, CA, TX, GA, TN, NC and PA.

Lot / Code Info

Femoral Tray - Model 0150ARR00001, Lot Number/Quantity: 0609038 (6), 06051107 (4), 01010806 (4), 09010108 (2), 06040707-01 (1), 01010107 (3), 01010607 (1), 01010106 (1) and 06040707 (2) and    Tibial Tray - Model 0150ARR00002, Lot Number/Quantity: 06060308 (6), 06061107 (7), 01020866 (2), 01010507 (1), 01020806 (1), 01020107 (3), 09081107 (2) and 01020607 (2).

Root Cause

Component design/selection

Action Taken

A MAKO employee will visually verify product at each account for any product affected by this recall. In addition, a recall letter dated July 1, 2008 was mailed to users (surgeon and owner of system, e.g. hospital). Customers were instructed to complete the enclosed Acknowledgement form and return it to your Clinical Technical Specialist (CTS) or in the enclosed pre-paid addressed envelope to the Mako Surgical Corporation. Please direct questions to your CTS or call 954-927-2044, extension 604.