Bio-Rad Variant II TURBO Hemoglobin Testing System with CDM v 4.0, model number 270-2600, Manufactured by Bio-Rad Laboratories, Hercules, CA Provides an integrated method for sample preparation, separation and the percent determination of percent hemoglobin A 1 c in EDTA human whole blood. This device only tests for A 1 c.

Company
Bio-Rad Laboratories Inc
Recall Initiated
October 17, 2008
Posted
March 24, 2009
Terminated
November 12, 2010
Recall Number
Z-1041-2009
Quantity
92 units
Firm Location
Hercules, CA

Reason for Recall

Sample results may be assigned to the wrong barcode number and tube position numbers, leading to discrepant sample results.

Distribution

Worldwide Distribution

Lot / Code Info

version 4.0 software

Root Cause

Pending

Action Taken

An Urgent Medical Device Correction letter was sent out on October 17, 2008 via fax and phone call. The letter recommended that customers review the Summary Report for duplicate tube position numbers or barcode IDs within the same rack. If there are questions or a need for assistance, customers are to contact their regional Bio-Rad office. Follow-ups will be made once root cause is identified.