Drill Bits Model #'s: DEP14043 MIC8053-019 RIC21-0438 RIC21-0439 RIC21-0441 RIC21-0442 RIC71173504 RIC71173505 RIC71631117 SMI21-0445 STR5800-4-125 STR60-13570 SYN03.010.060 SYN03.010.061 SYN03.010.101 SYN310.19 SYN310.24 SYN310.25 SYN310.288 SYN310.31 SYN310.35 SYN310.431 SYN310.44 SYN310.63 SYN310.65 SYN310.66 SYN310.67 SYN315.28 SYN315.40 SYN315.92 SYN317.861 SYN317.871 SYN356.982 SYN357.407 ZIM00-2318-020-00 and ZIM5056-146 An orthopedic manual surgical instrument is a nonpowered hand-held device intended for medical purposes to manipulate tissue, or for use with other devices in orthopedic surgery. This generic type of device includes the cerclage applier, awl, bender, drill brace, broach, burr, corkscrew, countersink, pin crimper, wire cutter, prosthesis driver, extractor, file, fork, needle holder, impactor, bending or contouring instrument, compression instrument, passer, socket positioner, probe, femoral neck punch, socket pusher, reamer, rongeur, scissors, screwdriver, bone skid, staple driver, bone screw starter, surgical stripper, tamp, bone tap, trephine, wire twister, and wrench.

This recall has been terminated (originally issued January 13, 2009).

Company
Sterilmed Inc
Recall Initiated
August 8, 2008
Posted
January 13, 2009
Terminated
April 26, 2011
Recall Number
Z-0548-2009
Quantity
99
Firm Location
Osseo, MN
Official Source
View on FDA website ↗

Reason for Recall

SteriMed, Inc is conducting a product removal of a small group of devices individually packaged and sealed within a pouch. The removal is being conducted because the possibility exists that some of the pouches were not properly sealed on one end. A breach in packaging seal or a failure in packaging integrity has the possibility of risk to the patient in terms of transmitting organisms capable

Distribution

AL, AZ, CA, CO, CT, FL, GA, IA, ID, IL, IN, KS, KY, MA, MI, MN, MO, MS, MT, NC, ND, NE, NJ, NY, OH, OK, OR, PA, SC, SD, TN, TX, UT, WA, WI, and WV OUS: CANADA

Root Cause

Packaging process control

Action Taken

Consignees were notified of the recall via verbal communication in person by on-site technician and or their local sales representative on 8/8/08. Others were contacted via fax or e-mail followed by on-site visits 8/11/08. SteriMed also provided consignees (Materials Manager/Risk Manager) with a "Urgent Product Removal" recall letter (August 14, 2008) in person informing them of the potential of receiving affected product. Included in the Product Removal Notification was a Customer Reply Form to inventory quantity of product received and quantity returned. Contact SterilMed, Inc. at 1-888-541-0078 for assistance.