Version 12 software program written for the Bio-Tek ELx800 automated microtiter plate reader with use of PRO-Trac Tacrolimus ELISA kit

Company
Diasorin Inc.
Recall Initiated
April 12, 2004
Posted
April 13, 2005
Terminated
July 22, 2006
Recall Number
Z-0702-05
Firm Location
Stillwater, MN

Reason for Recall

The software program written for use in the automated microtiter plate reader (Bio-Tek ELx800) for use with the Pro-Trac II Tacrolimus ELISA kit was found to be faulty. Assay software program did not include the Calibrator Zero OD specification (STDO>=1.500) as specified in the product insert.

Distribution

CA, Washington DC, FL, KY & Canada

Lot / Code Info

version 12

Root Cause

Other

Action Taken

Customers were sent notification letters and effectiveness check form. recommended users manually veify results using product insert specifications.