NAMIC Custom Angiographic Kit, Left Heart Kit. - . Model #: H7496019054110. Catalog #601905411. Firm on kit label: Manufactured by: Boston Scientific Corporation/NAMIC Technology Center, Glens Falls, NY 12801.
- Company
- Boston Scientific Corp
- Recall Initiated
- October 21, 2004
- Posted
- December 9, 2004
- Terminated
- October 13, 2005
- Recall Number
- Z-0303-05
- Quantity
- 24 kits
- Firm Location
- Tech Park Glens Falls, NY
Reason for Recall
The kits contains defective Magellan Monoject Safety Needles. The needle shaft may separate from the hub. Tyco Healthcare has recalled the needles.
Distribution
2 customers in IL and NV.
Lot / Code Info
988046
Root Cause
Other
Action Taken
Two customers contacted via phone on 10/21/04 and 10/22/04, followed by letters dated 11/17/2004. Customers instructed to discontinue use and return product. Distributor instructed to notify customers.