Bio-Rad VARIANT II Hemoglobin A1c Program Recorder Pack with Calibrator/Diluent Set, Model Number 270-2101NU Manufactured by Bio-Rad Laboratories, Hercules, CA

Company
Bio-Rad Laboratories Inc
Recall Initiated
June 15, 2009
Posted
March 5, 2010
Terminated
February 2, 2011
Recall Number
Z-1013-2010
Quantity
613 units
Firm Location
Hercules, CA

Reason for Recall

Calibration Failure-- due to high intercept values, the device fails calibration and users are unable to run samples.

Distribution

Product was distributed to 127 customers throughout the US and internationally to Argentina, Australia, Canada, Chile, Czech Republic, France, Germany, Hong Kong, Hungary, India, Italy, New Zealand, Poland, Singapore, Sweden, Taiwan, and the United Kingdom

Lot / Code Info

Medical device, intended for percent determination of hemoglobin A1c in human whole blood using ion-exchange high-performance liquid chromatography, intended for professional use only.

Root Cause

Other

Action Taken

Recall initiated June 15, 2009, with notification initially conducted by phone with Fax Notification follow up domestically. Each subsidiary will be given an urgent notice, subsidiary response form, customer notification, customer acknowledgement form. Each subsidiary is to contact customers and translate notification and response form as needed. Contact the recalling firm for additional information.