RecallDepth

American Medical Systems, Inc., (AMS) GreenLight HPS Fiber Optic (ADDStat) Part Number 0010-2090B. Fiber optic laser for treatment of BPH (Benign Prostatic Hyperplasia). AMS Innovative Center, San Jose, CA 95134

This recall has been terminated (originally issued February 26, 2008).

Company
Ams Innovative Center San Jose
Recall Initiated
June 18, 2007
Posted
February 26, 2008
Terminated
March 23, 2010
Recall Number
Z-0131-2008
Quantity
611 units
Firm Location
San Jose, CA
Official Source
View on FDA website ↗

Reason for Recall

Control knob break: The knob used to finely control the direction of the laser energy may break.

Distribution

Worldwide Distribution - USA and Singapore.

Lot / Code Info

Part Number: 0010-2090B, Lot Numbers 705 N and 706 E

Root Cause

Process design

Action Taken

American Medical Systems, Inc. (AMS) issued Important: Medical Device Recall AMS Greenlight ADDStat Fibers notification letters to customers who have purchased the product on June 28, 2007. The customer letter requests recipients that the product use be discontinued, and the fibers be returned. The firm will send replacements.