Datascope CS 100 Intra-Aortic Balloon Pump and Datascope CS 300 Intra-Aortic Balloon Pump, Datascope Corp. Mahwah, NJ 07430

This recall has been terminated (originally issued January 31, 2008).

Company
Datascope Corp.
Recall Initiated
November 19, 2007
Posted
January 31, 2008
Terminated
May 20, 2008
Recall Number
Z-0567-2008
Quantity
70 units US
Firm Location
Mahwah, NJ
Official Source
View on FDA website ↗

Reason for Recall

Defective Circuit Board -- A defect in the Printed Circuit Board's (PCB) in the CS 100 and CS300 Intra-Aortic Balloon Pump may cause the pump to exhibit intermittent malfunctions, which include failure to start-up or reset of the display screen during therapy.

Distribution

Worldwide Distribution

Lot / Code Info

Serial Numbers: S805991-C2 S805993-C2 SA02504-C4 S806332-G2 S82934-H9 SA03941-F5 S82416-D9 S805969-C2 S806909-B3 S806924-B3 SA03395-A5 SA03407-A5 SI101373-H7 S805404-H1 S81758-K8 S807356-F3 S807013-C3 S806289-F2 S806308-G2 S806282-F2 S806306-G2 S806284-F2 S806292-F2 S806297-G2 SA04871-A6 S806286-F2 S806926-B3 S804246-J0 S805797-A2 S804934-D1  S806155-E2 SI101367-H7 SI100417-H7 S807506-J3 S807042-C3 S81471-I8  SA02801-G4  S8233778-E0 S82518-E9 S803467-B0 SA03469-B5 S806229-F2 S804255-J0 S806193-E2 S806025-D2 S806012-D2 S806019-D2 S806045-D2 S806017-D2 S806024-D2 S805992-C2 S806009-D2 S805581-I1 S82350-C9 S804783-B1 S804793-B1 S804799-B1 S806859-A3 SA01430-A4 SA01152-I3 SA01421-A4 SA01412-A4 S803857-F0 S803775-E0 SA02521-D4 SI01410-E7  S807446-G3 S807470-H3 S806225-F2  and S806259-F2  (These serial numbers do not differentiate between CS100 and CS300, just the board contained within.)

Root Cause

Other

Action Taken

Datascope sent a Product Field Correction notification letters to Hospital Administrators, via certified mail, on November 19, 2007. The letter informed them of a defect in the circuit board. It described the possible malfunction and provided instructions should malfunctioning occur. Datascope will be replacing the defective circuit boards in all affected IABP units. A Service Representative will be contacting the hospital to arrange for the replacement of the defective board.