GE 4 inch Anterior Cervical Post, GE Part Number 1006385 or 1006385-NAV, GE Healthcare OEC Medical Systems (an accessory used with the InstaTrak 3500 plus and 9800 C-Arm Navigation systems)

This recall has been terminated (originally issued October 3, 2006).

Company
Ge Oec Medical Systems, Inc
Recall Initiated
May 31, 2006
Posted
October 3, 2006
Terminated
July 27, 2011
Recall Number
Z-0001-2007
Quantity
143 units
Firm Location
Salt Lake City, UT
Official Source
View on FDA website ↗

Reason for Recall

Weld defect- the weld between the post body and the screw was not properly formed and can fail, even when minimal force is used.

Distribution

Nationwide, including one VA facility in PA. No military distribution. Foreign distribution to Canada, China, Italy, Japan, Spain.

Lot / Code Info

ALL LOT NUMBERS -GE Part number 1006385 or 1006385-NAV

Root Cause

Process control

Action Taken

Consignees were notified by letter on 05/31/2006