Skytron General Purpose Surgical Table; Model 6600, Dist. by Skytron, Grand Rapids, MI.
- Company
- Skytron, Div. The Kmw Group, Inc
- Recall Initiated
- August 4, 2005
- Posted
- September 13, 2006
- Terminated
- May 1, 2008
- Recall Number
- Z-1496-06
- Quantity
- 160 total
- Firm Location
- Grand Rapids, MI
Reason for Recall
The release levers can inadvertently release under load while the leg section is positioned at a ninety degree angle causing the leg section to drop.
Distribution
Nationwide.
Lot / Code Info
All units. Serial numbers 6600-2D-001 through 6600-2L-102.
Root Cause
Other
Action Taken
A service bulletin was issued to distributors on 8/12/05 instructing them to identify customers and to retrofit these tables.