TERUMO KHURI MYOCARDIAL PH MONITORING SYSTEM, CATALOG NUMBER 7205

This recall has been terminated (originally issued September 29, 2006).

Company
Terumo Cardiovascular Systems Corp
Recall Initiated
June 23, 2006
Posted
September 29, 2006
Terminated
March 22, 2012
Recall Number
Z-1517-06
Quantity
175
Firm Location
Tustin, CA
Official Source
View on FDA website ↗

Reason for Recall

Lack of Assurance of Sterility: these specific lots of sensors lack supporting documentation for sterilization assurance.

Distribution

Nationwide in CA, IL, MA, MD, NC, & OK; Internationally in Belgium & Kuwait

Lot / Code Info

Lots Affected: 2006-O6AD, 2006-O8AD, 2006-O8AP,  6O8AP6O8AD

Root Cause

Other

Action Taken

The affected lots were identified as those lots manufactured in the first quarter of the calendar year 2006. The distribution of affected lots was traced. A letter and a customer response form were developed and each of the five US customers were sent the recall letter via Federal Express. The two international affiliates were contacted and instructed to contact their customers in a manner similar to one used by Terumo CVS in the US. Reconciliation will involve confirmation that the response form for each of the customers and affiliates has been completed and returned to Terumo CVS.