Prima III TotalSENSE Torso Coil (Body Array Coil), labeled as Manufactured by: USA Instruments, Inc. This device is a component of the Magnetom TRIO 3.0T MRI Systems, manufactured by Siemens AG Medical Solutions.

Company
Usa Instruments Inc.
Recall Initiated
March 19, 2004
Posted
July 20, 2004
Terminated
March 1, 2012
Recall Number
Z-0849-04
Quantity
79
Firm Location
Aurora, OH

Reason for Recall

A defective capacitor in the Body Array Coil may result in unstable, unreliable performance, with excessive heating of patient's body.

Distribution

The device was shipped to domestic consignees located CT, NY, PA, NC, GA, FL, OH, IN, MN, IL, UT, CA, OR; and foreign consignees in Austria, Denmark, Spain, Hong Kong, China, Netherlands, Japan, Taiwan, Korea, Australia, Turkey, Brazil, Germany.

Lot / Code Info

USAI P/N: 100352, USAI S/Ns: 310 to 388 (Rev 1 and 2); Siemens P/N: 7578789, Siemens S/Ns: 1010 to 1088.

Root Cause

Other

Action Taken

The firm notified the regional Siemens Service Centers by letter on 3/19/2004. The affected coils will be returned to the recalling firm for modification.