RecallDepth

BD Syringe Tip Cap Bulk Sterile Convenience Pak-1000 units/shelf carton. Used to seal syringes and maintain container closure integrity throughout the shelf life of the BD WWD syringe. Catalog Number: 305822

Company
Becton Dickinson & Company
Recall Initiated
September 27, 2021
Posted
October 27, 2021
Terminated
May 4, 2023
Recall Number
Z-0179-2022
Quantity
8,000 units
Firm Location
Franklin Lakes, NJ

Reason for Recall

Holes in product packaging impacts package integrity and may compromise sterility.

Distribution

TX

Lot / Code Info

Lot Number: 1111347 Exp. Date: 4/30/2026 UDI: 10382903058225

Root Cause

Equipment maintenance

Action Taken

BD notified consignees Urgent Medical DevIce Recall Letter dated 9/7/21 via FedEx and may also be contacted via e-mail (when available). Letter states reason for recall, health risk and action to take: 1. Immediately review your inventory for the specific Catalog (Ref) and Lot number listed above. Destroy all product subject to the recall following your institution s process for destruction. 2. Share this recall notification with all users within your facility network of the product to ensure they are also aware of this recall. 3. Complete the attached Customer Response Form and return to the BD contact noted on the form whether or not you have any of the impacted material so that BD may acknowledge your receipt of this notification per FDA requirements and subsequently process any applicable product replacement. 4. Report any adverse health consequences experienced with the use of this product to BD. BD will provide replacement product for all discarded inventory. Contact Information: If you require further assistance, please contact: BD Contact Contact Information North American Regional Complaint Center 1-84BD-LIFE (1-844-823-5433) Say Recall when prompted Mon Fri 8:00am and 5:00pm CT

More recalls by Becton Dickinson & Company