Product Name: Lumipulse -Amyloid Plasma Controls Model/Catalog Number: 81559 Software Version: Not Applicable Product Description: Lumipulse -Amyloid Plasma Controls: This quality control product can be used as an objective judgement of the laboratory s procedures and personnel techniques when analyzing -amyloid1-42 in K2EDTA plasma on the LUMIPULSE G System. It is a valuable tool to assess good laboratory practices. Two levels of control are available to compare observations with expected ranges therefore assuring consistent performance of the testing system within the clinical range. Lumipulse -Amyloid Plasma Controls: Liquid (Frozen), 2 2 Concentrations L1 Level 1 (2 1.5 mL) L2 Level 2 (2 1.5 mL) Contains Tris buffer with protein (bovine) and chemical stabilizers. Preservative: 0.1% ProClin 950 Component: Yes. Lumipulse G pTau217/B-Amyloid 1-42 Plasma Ratio
- Company
- Fujirebio Diagnostics, Inc.
- Recall Initiated
- December 11, 2025
- Posted
- February 5, 2026
- Recall Number
- Z-1307-2026
- Quantity
- 179 units
- Firm Location
- Malvern, PA
Reason for Recall
Customers may experience inaccurate test results by observing a higher frequency of classification in the indeterminate and positive zone and/or low specificity when compared back to FDA-cleared CSF ratio/PET imaging, due to falsely elevated positive ratio or falsely elevated indeterminate ratio results causing an incorrect classification of a patient as having amyloid pathology associated with Alzheimer s disease (positive result) and/or requiring more clinical testing of that patient (indeterminate result).
Distribution
US Nationwide distribution in the states of Arizona, California, Florida, Indiana, Kansas, Kentucky, Maryland, Massachusetts, Minnesota, New Jersey, Pennsylvania, Texas, Washington.
Lot / Code Info
Model/Catalog Number: 81559; UDI-DI: 05414470815593; Lot Number: T8B6116U;
Root Cause
Under Investigation by firm
Action Taken
On or about December 11, 2025 URGENT PRODUCT CORRECTION NOTIFICATION letters were sent to customers. Actions to be taken: - Immediately discontinue using the associated Lots listed above for the Lumipulse G pTau217/-Amyloid 1-42 Plasma Ratio. - Review this letter with your Medical Director to determine the appropriate course of action, including for any previously generated results, if applicable. - Notify your ordering physician or health care provider as appropriate. - The test results must be interpreted in conjunction with other diagnostic tools and clinical information. - Please retain this letter with your laboratory records and forward this leer to those who may have received this product.