Brand Name: Alcon Laboratories, Inc. Product Name: Accessory devices for Unity and Constellation surgical consoles Model/Catalog Number: Many - See 'Attachment 1 - Affected Product List' 8065000393 UNITY HP VISCOUS FLUID CONTROL VFC PACK Software Version: N/A Product Description: These products are standalone accessory devices meant for vitrectomy surgical procedures. They interface with a corresponding Alcon surgical console. Component: N/A
- Company
- Alcon Research Llc
- Recall Initiated
- August 5, 2025
- Posted
- October 24, 2025
- Recall Number
- Z-0306-2026
- Quantity
- 6
- Firm Location
- Worth, TX
Reason for Recall
Due to incomplete seals in the pouch which provide the sterile barrier.
Distribution
Worldwide distribution: US (nationwide): AK, AL, AR, AZ, CA, CO, CT, DC, DE, FL, GA, HI, IA, IL, IN, KS, KY, LA, MA, MD, ME, MI, MN, MO, MS, MT, NC, ND, NE, NH, NJ, NM, NY, OH, OK, OR, PA, RI, SC,SD,TN, TX, VA, WA, WI, and WV; and OUS (foreign) to countries of: Argentina, Australia, Austria, Azerbaijan, Bahrain, Bangladesh, Belgium, Brazil, Brunei, Canada, Chile, China, Colombia, Croatia, Czechia, Denmark, Finland, France, Germany, Greece, Guadeloupe, Hong Kong, Hungary, Iceland, India, Ireland, Israel, Italy, Japan, Korea, Kosovo, Luxembourg, Malaysia, Mexico, Moldova, Montenegro, Morocco, Netherlands, New Zealand, Pakistan, Peru, Philippines, Poland, Portugal, Qatar, Reunion, Russian Federation, Saudi Arabia, Serbia, Slovakia, South Africa, Spain, Sri Lanka, Sweden, Switzerland, Taiwan, Turkey, United Arab Emirates, and United Kingdom,
Lot / Code Info
Catalog Number: 8065000393 UDI-DI code: 00380650003935 Lot/Batch Number: 17CDKX
Root Cause
Process control
Action Taken
On 08/05/2025, the firm sent via FedEx Overnight an "URGENT: MEDICAL DEVICE RECALL" Letter to customers informing them that for the Alcon Standalone Vitrectomy Consumables, there is potential for some pouches within impacted lots to have an incomplete seal. Customers are instructed to: locate and dispose of any affected lots of Alcon Standalone Vitrectomy Consumables remaining in your inventory. To comply with this Medical Device Recall and request the replacement of any unused product, please take the following steps: 1. Review inventory to determine if there are any unused affected product within their facility. See table on page 1 for affected Standalone Vitrectomy Consumable lots shipped to your location. 2. Segregate and dispose of any unused affected product from their inventory. 3. Call Alcon Customer Service to arrange for replacement of the affected inventory of Alcon Standalone Vitrectomy Consumables. 4. Respond to Alcon indicating understanding of these instructions even if they have zero (0) units remaining in inventory by completing the online response form at https://qrco.de/Vit-Seal or by completing and returning the attached Response Form and returning to Alcon via email: Market.Actions@Alcon.com or fax: 817-302-4337. 5. Forward this notification to all departments within their organization who may be in possession of this affected product; and any other organization to which this product may have been transferred. For questions contact, Alcon Customer Service at 1-800-862-5266 or an Alcon Sales Representative.