BEAR Implant (Bridge-Enhanced ACL Restoration); Model/Catalog: 1000;

Company
Miach Orthopaedics
Recall Initiated
February 2, 2023
Posted
November 7, 2025
Recall Number
Z-0479-2026
Quantity
167 units
Firm Location
Westborough, MA

Reason for Recall

The field action was initiated in February of 2023 and was limited to the affected lot from sales force trunk stock only.

Distribution

US Nationwide distribution.

Lot / Code Info

Model/Catalog: 1000; UDI-DI: 00860002987804; Lot number: 7008409;

Root Cause

Under Investigation by firm

Action Taken

On January 30, 2023 RE: BEAR Implant Hydration memo emails were sent to sales staff. If you observe an implant that hydrates too quickly (e.g., less than 30 seconds) and appears that it may not remain intact upon insertion into the knee, do not attempt to implant the device. We would recommend you always have a backup device available to address such circumstances. We plan to revise the Tips and Pearls document, ML-1027, to provide further instructions/details on how to properly hydrate the BEAR device. In the meantime, one way to avoid shortened hydration time is to consider hydrating the dense side of the implant first and continue toward the porous side . To date, all reports of this observation have involved Lot # 7008409. As a precaution and to ensure customer satisfaction, we are asking that you return all devices from this lot that are currently in your trunk stock (in your possession). Chris Brown will work with you to coordinate returns as well as replacing these trunk stock devices. Please contact me (mohara@miachortho.com) or Arjun (aishwar@miachortho.com) if you have questions.

More recalls by Miach Orthopaedics

View all recalls by this company →