Covidien Signia" Small Diameter Curved Tip Intelligent Reload (45mm Vascular/Thin); BOX LOBECTOMIA VATS 3D; CFN: BOX06170V4;
This recall is currently active, issued November 7, 2025.
- Recall Initiated
- September 25, 2025
- Posted
- November 7, 2025
- Recall Number
- Z-0486-2026
- Quantity
- 2 units (OUS only)
- Firm Location
- North Haven, CT
- Official Source
- View on FDA website ↗
Reason for Recall
A manufacturing issue could lead to a malformed cover tube press which can lead to internal components' movement during firing - leading to the potential for uncontrolled motion.
Distribution
Worldwide distribution - US Nationwide and the countries of Austria, Belgium, Chile, Estonia, Finland, France, Germany, Greece, Hungary, Israel, Italy, Japan, Kuwait, Malaysia, Netherlands, Poland, Portugal, Romania, Saudi Arabia, Slovakia, Spain, United Kingdom.
Lot / Code Info
CFN: BOX06170V4; UDI-DI: 00763000988975; Lot numbers: 0231961695;
Root Cause
Process control
Action Taken
On September 25, 2025 URGENT: MEDICAL DEVICE RECALL letters were sent to customers. Customer actions: Identify and quarantine all unused impacted product listed. Return all quarantined devices to Medtronic. If purchased from a distributor, contact your distributor directly to arrange for the return of the devices back to your distributor. Please complete and return the enclosed Customer Confirmation Form. Pass on this notice to all those who need to be aware within your organization or to any organization where the listed affected lots may have been transferred or distributed.