BioPro Endo Head (Metal uni-polar, 38-55) Product Number/Description 10179 ENDO MD 38 10180 ENDO MD 41 10181 ENDO MD 43 10182 ENDO MD 45 10183 ENDO MD 47 10184 ENDO MD 49 10185 ENDO MD 51 10186 ENDO MD 53 10187 ENDO MD 55 10188 ENDO SH 38 10189 ENDO SH 41 10190 ENDO SH 43 10191 ENDO SH 45 10192 ENDO SH 47 10193 ENDO SH 49 10194 ENDO SH 51 10195 ENDO SH 53 10196 ENDO SH 55
- Company
- Biopro, Inc.
- Recall Initiated
- September 9, 2025
- Posted
- October 14, 2025
- Recall Number
- Z-0149-2026
- Quantity
- 147 total
- Firm Location
- Port Huron, MI
Reason for Recall
Potential packaging issue that could compromise the sterile barrier. Defect may result in loss of sterility of the device, presenting potential risk of infection if device is implanted.
Distribution
US Domestic distribution to Texas and Michigan.
Lot / Code Info
Product ID/UDI-DI 10179 M20910179 10180 M20910180 10181 M20910181 10182 M20910182 10183 M20910183 10184 M20910184 10185 M20910185 10186 M20910186 10187 M20910187 10188 M20910188 10189 M20910189 10190 M20910190 10191 M20910191 10192 M20910192 10193 M20910193 10194 M20910194 10195 M20910195 10196 M20910196 All lots, all serial numbers
Root Cause
Package design/selection
Action Taken
Consignees were notified on September 9, 2025, via phone, followed by formal letters with instructions for returning the inventory to BioPro.
More recalls by Biopro, Inc.
BioPro Femoral Head (Ceramic heads 28-30)
Product ID/Description
19023 HEAD FEMORAL CERAMIC 28MM -3...
Oct 14, 2025
BioPro Femoral Head
Product ID/Description
14089 HEAD FEMORAL 28 SH LT
14090 HEAD FEMORAL 28 MD LT...
Oct 14, 2025
BioPro Bipolar Head
Product ID/Description
18130 BIPOLAR HEAD 38MM
18131 BIPOLAR HEAD 39MM
18132 BI...
Oct 14, 2025