Sprinter Cart/Sprinter Cart XL Infusion Pole. Component used to hold 2 x 2 kg infusion bags.

Company
Maquet Cardiopulmonary Gmbh Kehler Str. 31 Rastatt Germany
Recall Initiated
June 27, 2025
Posted
July 29, 2025
Recall Number
Z-2193-2025
Quantity
7,754 units

Reason for Recall

Device does not comply with the labeling requirements of international standard IEC 60601-1.

Distribution

Worldwide distribution - US Nationwide and the countries of Australia, New Zealand, Canada, Hong Kong, India, Israel, Japan, Singapore, South Korea, Taiwan, Thailand, Austria, Belgium, Bulgaria, Croatia, Cyprus, Czech Republic, Denmark, Finland, France, Germany, Greece, Hungary, Ireland, Italy, Lithuania, Luxembourg, Netherlands, Norway, Poland, Portugal, Serbia, Slovakia, Slovenia, Spain, Sweden, Turkey, Reunion, Romania, Russian Federation, Switzerland, Ukraine, United Kingdom, Chile, Colombia, Costa Rica, Dominican Republic, Ecuador, Mexico, Paraguay, Peru, Uruguay, Bahrain, Egypt, Iran, Iraq, Jordan, Kuwait, Lebanon, Oman, Pakistan, Qatar, Saudi Arabia, South Africa, United Arab Emirates, Yemen.

Lot / Code Info

Model No. 701033599; UDI: 4037691257860.

Root Cause

Under Investigation by firm

Action Taken

An URGENT MEDICAL DEVICE CORRECTION NOTICE dated June 2025 was mailed to consignees. The notification informs consignees that Getinge will be sending a field service technician to visit and verify if a safety sign is present on affected devices. Consignees are asked to share the recall notification with users of affected devices or to facilities to which product was further distributed for their awareness. Consignees are to complete and return the provided response form to recallresponses.qrc@getinge.com. Consignees with any questions are to direct them to Technical Support at 1-877-943-8872 (options 4, 2, 1) or by email at techsupport.us@getinge.com.

More recalls by Maquet Cardiopulmonary Gmbh Kehler Str. 31 Rastatt Germany

View all recalls by this company →