RecallDepth

Apex Knee System. Model Numbers: KC-50090, KC-50097, KC-50110, KC-50117, KC-50130, KC-50135, KC-50137, KC-50150, KC-50155, KC-50157, KC-50170, KC-50175, KC-50177, KC-50190, KC-50197, KC-52104, KC-52554, KC-52574, KC-52774, KC-55223, KC-55243

This recall is currently active, issued July 2, 2025. It was issued by Corin Medical, Ltd. Unit 1-57-8 Corinium Centre Cirencester United Kingdom.

Company
Corin Medical, Ltd. Unit 1-57-8 Corinium Centre Cirencester United Kingdom
Recall Initiated
May 23, 2025
Posted
July 2, 2025
Recall Number
Z-2069-2025
Quantity
341 units
Official Source
View on FDA website ↗

Reason for Recall

Investigation has identified that the age of the nylon packaging material has caused a loss of integrity of some pouches.

Distribution

US Nationwide distribution in the states of CA, CO, FL, IL, OK, TX, UT.

Lot / Code Info

Model Number (Serial Numbers): KC-50090(580854, 582278), KC-50097(528456, 528457), KC-50110(531527, 532301, 586311), KC-50117(552924), KC-50130(528458, 548885, 548886, 548887, 548888, 573259), KC-50135(529093), KC-50137(527387, 528459, 555044, 568751), KC-50150(529072, 529094, 565748, 584973, 584974), KC-50155(586369), KC-50157(528460, 528461, 545943, 565831, 565832, 567009, 572398, 587261, 587262), KC-50170(528462, 529076, 550671, 585011), KC-50175(552532), KC-50177(529073, 588382), KC-50190(529074), KC-50197(548468, 548469, 548470), KC-52104(584975), KC-52554(589609), KC-52574(586115, 586117), KC-52774(584976), KC-55223(580117), KC-55243(580855).

Root Cause

Nonconforming Material/Component

Action Taken

Corin Group notified US agents informally via email on 05/23/2025 and informed them of an investigation and affected inventory to quarantine. A formal notification was sent to consignees on about 06/02/2025. Consignees were instructed to identify any affected units on hand and quarantine them, return them to Corin and notify all affected parties who handled the product. Consignees were also requested to complete and return the acknowledgement form. If product was further distributed, Corin requested contact information to contact those customers.