GE Vivid Ultrasound products, Model Numbers: a) Vivid S70, H45041SW

Company
Ge Vingmed Ultrasound As Strandpromenaden 45 Horten Norway
Recall Initiated
May 2, 2024
Posted
July 5, 2024
Recall Number
Z-2317-2024
Quantity
1 unit

Reason for Recall

GE HealthCare has become aware of an issue with certain Vivid ultrasound systems where, due to a slow DICOM Worklist response time, incorrect patient name and demographics can be displayed on the ultrasound scanner screen rather than the intended patient selected by the user.

Distribution

US: IL, MA, NM, NY, OK,

Lot / Code Info

a) H45041SW, UDI/DI *, Serial Numbers: 001423S70

Root Cause

Device Design

Action Taken

GE Healthcare issued an URGENT MEDICAL DEVICE CORRECTION notice to its consignees on 05/02/2024 via letter delivered by traceable means. The notice explained the problem with the device, potential safety issue, and requested the following: "Actions to be taken by Customer/User: You can continue to use your device in accordance with the User Manuals, specifically: 1. After selecting a patient in the DICOM Modality Worklist, make sure to review and confirm that the patient demographics shown on the ultrasound scanner screen during the exam are from the intended patient. 2. In the scenarios where the incorrect and/or actual patient already exists in the patient archive, a Patient Match dialog can be shown to inform you about patient information inconsistency. Make sure to review the patient demographics properly. 3. To reduce the probability of occurrence of the issue, reduce the size of the DICOM Worklist search by: a) Setting Max. Results to 75 in the Worklist dataflow Config dialog (available by selecting Config | Connectivity | Dataflow | Worklist* | Inputs | DicomWorklist | Properties) and/or b) Configuring the DICOM Worklist server to automatically delete Scheduled Procedure Steps that have been performed on the ultrasound scanner. If this option is not available on your DICOM Worklist server, make sure to manually clear all Scheduled Procedure Steps that have been performed from the DICOM Worklist on a regular basis (e.g., every day). *The actual instructions on how to perform this step depends on which DICOM Worklist server is used in your department. Consult your IT department to find out how to perform the step. Please ensure all potential users in your facility are made aware of this safety notification and the recommended actions." "GE HealthCare will correct all affected products at no cost to you. A GE HealthCare representative will contact you to arrange for the correction."

More recalls by Ge Vingmed Ultrasound As Strandpromenaden 45 Horten Norway

View all recalls by this company →