MobiCT-32 AIRO Mobile CT Scanner, Cat. No. MobiCT-32, mobile computed tomography (CT) system, software version 2.1.1

Company
Mobius Imaging, Llc
Recall Initiated
November 18, 2022
Posted
December 6, 2022
Terminated
February 26, 2025
Recall Number
Z-0344-2023
Quantity
3
Firm Location
Shirley, MA

Reason for Recall

Stryker received a report of a battery fire involving an Airo TruCT unit previously subjected to bypass charging. The firm is notifying customers with units whose battery monitoring systems were bypassed to charge the batteries; the batteries in these units require replacing.

Distribution

Domestic distribution to consignees in Colorado, Tennessee, and Utah.

Lot / Code Info

UDI-DI 00869346000200 Serial Numbers AIRO-0124, AIRO-0145, AIRO-0288

Root Cause

Other

Action Taken

On November 18, 2022, the firm notified affected customers via Urgent Medical Device Correction letters. Customers were informed that a service representative will contact their facility to schedule a time to replace the batteries in the affected devices. The customer does not need to take additional action while waiting for the battery preplacement. If you have any questions or concerns regarding this update, please contact shirleytechnicalsupport@stryker.com or call Stryker Technical Support at 978 386 9619.